Vulnerability Name:

CVE-2017-12722

Assigned:2017-08-09
Published:2017-09-07
Updated:2018-02-15
Summary:An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump reads memory out of bounds, causing the communications module to crash. Smiths Medical assesses that the crash of the communications module would not impact the operation of the therapeutic module.
CVSS v3 Severity:5.3 Medium (CCN CVSS v3 Vector: CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:L)
4.7 Medium (CCN Temporal CVSS v3 Vector: CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:L/E:U/RL:U/RC:R)
Exploitability Metrics:Attack Vector (AV): Network
Attack Complexity (AC): Low
Privileges Required (PR): None
User Interaction (UI): None
Scope:Scope (S): Unchanged
Impact Metrics:Confidentiality (C): None
Integrity (I): None
Availibility (A): Low
CVSS v2 Severity:5.0 Medium (CCN CVSS v2 Vector: AV:N/AC:L/Au:N/C:N/I:N/A:P)
Exploitability Metrics:Access Vector (AV): Network
Access Complexity (AC): Low
Athentication (Au): None
Impact Metrics:Confidentiality (C): None
Integrity (I): None
Availibility (A): Partial
References:Source: BID
Type: UNKNOWN
100665

Source: BID
Type: UNKNOWN
101252

Source: XF
Type: UNKNOWN
smithsmedical-cve201712722-dos(131639)

Source: MISC
Type: UNKNOWN
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A